Mainstream EtCO2 Sensor
Mainstream EtCO2 Sensor — view 2

Mainstream EtCO2 Sensor

Model: ET100

Product Description

Product Introduction

The ET100 Sensor is a highly integrated mainstream CO2 Sensor which developed by RSD medical, who had tens of years IR measurement instrument development experience. It’s the one-step capnography solutionfor patient monitoring system. Within a very compact size, it provides measurement of End-Tidal Carbon Dioxide(ETCO2), respiration rate, capnogram, which can cover most of requirement from the patient monitoring system. With a very simple RS232 interface and industry compatible protocol, it’s fairly easy to connect ET100 to any patient monitoring system and other medical treatment system where CO2 sensor is needed.

With the stability, flexibility and easy to use as the built in essence, ET100 is your ideal choice for your capnography subsystem.

Product Features

1.Real time CO2 monitoring and data upload

2.Dual wavelength , non-dispersive infrared(NDIR)

3.With fast electrical modulation MEMS IR source, moving part was avoid and long lifetime was achieved

4.Includes temperature control system, to prevent airway adapter condensing

5.Barometer pressure and Gas temperature compensation

6.Patent on calibration algorithm, no routine user calibration required

7.Small size and lightweight

SPECIFICATIONS

Transducer Type

Mainstream CO2 Sensor

Principle of Operation

Non-dispersive infrared(NDIR), dual wavelength, no moving parts

Initialization Time

Capnogram displayed in less than 5 seconds,full specifications within 2 minutes

CO2 Measurement Range

0 – 114 mmHg, 0 – 15 %, 0 – 15.2 kPa(760mmHg)

Rise Time

Less than 60 ms - Adult Reusable or Single-Patient-Use Airway Adapter

Less than 60 ms - Infant Reusable or Single-Patient-Use Airway Adapter

CO2 Resolution

0.1mmHg 0 to 59 mmHg

0.25 mmHg 60 to 114mmHg

CO2 Accuracy

0 – 40mmHg ±2 mmHg

41 – 76 mmHg ±5% of reading

77 – 114 mmHg ±8% of reading

Above 80 BPM ±12% of reading

CO2 Stability

Short term drift: Drift over four hours shall not exceed 1 mmHg maximum

Long term drift: Accuracy specification will be maintained over a 120-hour period

Sampling Frequency

100 Hz

Respiration Rate Range

2 to 150 Breaths Per Minute (BPM)

Respiration Rate Accuracy

±1 breath

Compensations

(Supplied)

Barometric pressure: 400 mmHg to 800 mmHg

Operator selectable O2, N2O, He and agent compensation

Calibration

No routine user calibration required. An airway adapter zero is required when changing

Airway Adapters

Single-patient-use or reusable, < 5 cc dead space (adult), < 1 cc dead space infant

Voltage Requirements

5.0 VDC ±5%

Power Consumption

1.2Watts typical (Steady State)

Up to 1.8 Watts maximum on power up (Warm up)

Interconnection

(The RS232 connector version is also provided)

Standard - Lemo Redel 8-pin plastic connector

Pin out:

1. VA 5.0VDC

2. Shield Shield

3. DGND Digital return

4. VSRC 5.0VDC

5. TxD Serial data from Sensor

6. RxD Serial data from Host

7. AGND Analog return

8. SYNC Unused

Data Interface

RS232 , bi-directional, 19200 baud rate, standard N-8-1.

Data Output

CO2 gas concentration(mmHg), End-tidal CO2, Inspired CO2, Respiratory Rate. Gas and barometric pressure compensated when supplied by host.

Temperature and Humidity

Operating: 0 to 45℃, 10 to 90% RH, non-condensing

Storage: -40 to 70℃, <90% RH, non-condensing

Water Resistance

IPX4 – Splash-proof (sensor head only)