Features:
1.Combine the Masimo technology and the Nellcor technology
2.With the advanced DSP algorithm, the measurement accuracy can be compare with
the word leading production in the movement and low perfusion environment.
3.Accurate measurement in 0.025% Low perfusion Index
4.Accurate measurement in all kind of extreme environment measurement such as
neonatal and shock patients
Measurement range:
SpO2 saturation: 0-100%
Pulse rate: 0-254BPM
Perfusion Index:0.025%-20%
Accuracy range:
SpO2 saturation: 70-100%
Pulse rate: 20-254BPM
Perfusion Index:0.025%-20%
Measurement Accuracy:
SpO2 saturation:
Adult (including pediatric ) + 2digit(70-100%) not defined (<70%)
Neonate + 3 digits not defined(<70%)
On motion condition + 3digit
Pulse rate:
Adult (including pediatric) + 3digits
Neonate + 3 digits
On motion condition + 3digit
Perfusion Index:
0.025%-20%
Resolution:
SpO2 saturation: 1%
Pulse rate: 1BPM
Numerical Calculation:
SpO2 saturation, pulse rate, signal strength, histogram ,volume recording wave,
Perfusion Index.
Mechanical Characters:
Width 38.3mm
Length 51mm
Height 14mm
Weight 11g
Assembly The module has 2 assembly holes
Electrical Characters:
Voltage 5V~6V DC
Static current <60mA
Current in the measurement <90mA